Test Overview
DCA Vantage- spectrophotometry
Afinion HbA1c is a fully automated boronate affinity assay for the determination of the percentage of hemoglobin A1c in human whole blood.
Hemoglobin A1c is proportional to the level of glucose in the blood over 2 to 3 months. Regular measurement of HbA1C can be used in the long-term clinical monitoring of diabetic patients. Improvement of metabolic control is indicated by a lowering of HbA1C values.
Point of Care DCA Hemoglobin A1C and Afinion 2 testing is available in certain ambulatory health centers and clinics and is performed by authorized operators in those areas.
Reference ranges are available in the DCA Vantage - Hemoglobin A1C and Afinion 2 test procedures located on the POC website: https://www.pathology.med.umich.edu/point-of-care
Reference ranges may change over time. Please refer to the original patient report when evaluating results.
* Reference ranges may change over time. Please refer to the original patient report when evaluating results.
The DCA Vantage will measure values between 2.5%-14%. Report out of range patient results as less than (<) 2.5% or greater than (>) 14.0%.
Specimens with hemoglobin values less than 7 g/dL should be tested with a different method.
The Afinion 2 will measure values between 4.0%-15%.The Afinion 2 has the following limitations:
1) Any cause of shortened erythrocyte life span will reduce exposure of erythrocytes to glucose, resulting in a decrease in HbA1c values, regardless of the method used. Caution should be used when interpreting the HbA1c results from patients with conditions such as hemolytic anemia or other hemolytic diseases, homozygous sickle cell trait, pregnancy, blood loss, polycythemia, iron deficiency etc.
2) Diluted samples cannot be used with Afinion HbA1c.
3) Coagulated or hemolyzed samples cannot be used with Afinion HbA1c. Samples with >14% (2000 mg/dL) hemolysis may return an information code.
4) If the sample has a hemoglobin value below 6.0 g/dL or above 20.0 g/dL, no test result will be reported, and an information code will be displayed.
Test Details
6 minutes
Specimen Requirements
Fingerstick: wipe away the first drop and use second drop for testing. Avoid excessive squeezing or milking. Afinion 2 does not require the first drop to be wiped away.
Venous: collect in lithium, sodium, or EDTA syringe/tube (Afinion Citrate also acceptable).
Sample must be tested within 5 minutes after collection in glass capillary.
Reject any sample that has remained in the glass capillary for longer than five minutes prior to analysis.
Additional Information
For more test specific information, please refer to the test procedure found on the Michigan Medicine POC website: https://www.pathology.med.umich.edu/point-of-care