Test Overview
Test Methodology
              Polymerase chain reaction
Test Usage
              Detection of Measles viral infection
Reference Range *
              No virus detected
* Reference ranges may change over time. Please refer to the original patient report when evaluating results.
Test Details
Days Set Up
              Monday - Friday (as needed)
                Analytic Time
              1 - 2 weeks
Soft Order Code
              MSLPC
                Synonyms
              Laboratory
              Sendout
                Reference Laboratory
              MDHHS
                Section
              Special Testing
                Specimen Requirements
Collection Instructions
              Collect specimen: Throat, oropharyngeal or nasopharyngeal with a Dacron synthetic swab and place into 3mL of VTM or M4 media. Send to Specimen Processing refrigerated.
Special Handling
              Send to MDHHS refrigerated. In case of delayed shipment, specimen may be frozen and sent
Normal Volume
              Dacron (synthetic) swab with aluminum or plastic shaft in 3 mL Viral Transport Medium (VTM) or M4 media
Minimum Volume
              Dacron (synthetic) swab with aluminum or plastic shaft in 3 mL Viral Transport Medium (VTM) or M4 media
Storage Temperature
              Refrigerated, frozen if shipment is delayed
                Rejection Criteria
              Swabs with calcium alginate or cotton tips with wooden shafts. or shipping the specimen at room temperature
                Additional Information
Test performed by MDHHS
Billing
NY State Approved
              No